How often do you need to change the internal desiccant on the FL1500?

Dokument-ID

Dokument-ID TE8822

Veröffentlichungsdatum

Veröffentlichungsdatum 20.06.2019
Frage
How often do you need to change the internal desiccant on the FL1500?
Zusammenfassung
Changing the internal desiccant on the FL1500.
Antwort
The internal desiccant life is highly dependent on the humidity exposure and continuity of the enclosure sealing during installation. It is best practice to proactively replace the desiccant when the enclosure main compartment is opened especially in high humidity locations. The desiccant can mitigate the enclosure volume at 60° C and 90% RH 3 times. The internal desiccant should be replaced or dried every third opening of the enclosure in a 60° C and 90% RH environment.

Replacement Desiccant can be ordered with Product # 8755500.