How do you connect a Sigma Rain Gauge to an FL900 and program it with Flo-Ware?

Dokument-ID

Dokument-ID TE8969

Veröffentlichungsdatum

Veröffentlichungsdatum 13.01.2022
Frage
How do you connect a Sigma Rain Gauge to an FL900 and program it with Flo-Ware?
Zusammenfassung
Connecting a Sigma Rain Gauge to an FL900 and programming it using Flo-Ware .
Antwort
  1. There is a Port on the FL900 labeled "Rain."  It is next to Port 4.  Connect the Rain Gauge cable to this Port.
  2. Connect your computer to the FL900
  3. Flo-Ware comes up on the "Programming Page." On the white rectangular box on the left side, you will see a list beginning with "Site Information."  The third item is "Datalog Setup."  Click on it.
    • If Flo-Ware is not on the "Programming Page," then go to the vertical tab on the far left of the grey box.  The first tab is "Programming."  Click on it.
  4. Skip the "Port Assignments" section.
  5. Below this is the "Select Datalog Channels" section.  To the left of "Logger" is a small box with a "+" in it.  Check on the "+".
  6. Under "Logger," "Power Supply" and "Rain Gauge" will appear.
  7. Click the box next to "Rain Gauge."  Now to the right you will see the "Rain" and three fields beneath it, which will allow you to select the following:
  8. Be sure to "Write to Logger" (button on lower right hand corner) so your changes will be written to the logger. 