How should the Lead Screws/Worm Gears on a Lachat ASX - series Sampler be cleaned?

Dokument-ID

Dokument-ID TE9179

Veröffentlichungsdatum

Veröffentlichungsdatum 13.12.2019
Frage
How should the Lead Screws/Worm Gears on a Lachat ASX - series Sampler be cleaned?
Zusammenfassung
Cleaning of the Lead Screws/Worm Gears on an ASX - series Sampler.
Antwort
The Lead Screws/Worm Gears on an ASX - series Sampler can be cleaned using alcohol and a dry cloth. One should continue to clean the Lead Screws until the cloth no longer picks up debris, even after moving the probe arm back and forth.

One should not apply silicone grease to the Lead Screws. Although the probe will move more smoothly, the grease will attract dust or other debris. Overtime, this can negatively impact functionality.