What can I use to clean Orbisphere Luminescent Dissolved Oxygen (LDO) series sensors such as the K1x00, M1100 & the 3100?

Dokument-ID

Dokument-ID TE9225

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
What can I use to clean Orbisphere Luminescent Dissolved Oxygen (LDO) series sensors such as the K1x00, M1100 & the 3100?
Zusammenfassung
Cleaning the Orbisphere LDO series sensors.
Antwort
Hach recommends using Ethanol (EtOH) to clean our LDO sensors. We have validated the cleaning at the factory with 96% EtOH, so you can use up to 100% EtOH without issue. Do not forget to dry the path with air or N2 after cleaning with EtOH.

Denatured alcohol which is a mixture of ethanol and methanol is an acceptable option also.

Use of rubbing alcohol or anything that contains Isopropyl alcohol is not recommended due to possible bleaching of the spot.