How does the Rapidly Settling Turbidity (RST) mode calculate results in TL23?

Dokument-ID

Dokument-ID TE7813

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
How does the Rapidly Settling Turbidity (RST) mode calculate results in TL23?
Zusammenfassung
Methodology of RST, Rapidly Settling Turbidity.
Antwort
The Rapidly Settling Turbidity (RST) mode calculates and continuously updates the turbidity reading of the sample to a confidence of 95%, based on the accumulated trend of the real time measured values. The RST mode is best used on samples that settle rapidly and continuously change in value. The reading is based on a correctly prepared sample that is homogeneous at the beginning of the reading. It is best applied to samples that are greater than 20 NTU. The sample must be mixed thoroughly by inversion immediately before inserting it into the meter. The target icon is shown on the display when the Rapidly Settling Turbidity is on. The Rapidly Settling Turbidity reads and calculates five readings while showing intermediate results.
 