What is the acceptable process pH range for the Nitratax sc sensor?

Dokument-ID

Dokument-ID TE9323

Veröffentlichungsdatum

Veröffentlichungsdatum 19.02.2018
Frage
What is the acceptable process pH range for the Nitratax sc sensor?
Zusammenfassung
Acceptable pH range for the Nitratax sc sensor
Antwort
The acceptable pH range for all Nitratax sc sensors is 4.5 pH to 9 pH.
 
If the sensor is installed outside of this pH range, there is a risk of a shorter lifespan on seals, wipers and/or sensor cable. Additionally, there could be a higher risk of leakage which could result in a total loss of the sensor due to it becoming filled with water. In some cases the risk can be mitigated by ordering service four times per year to check the sensor before damage is done. If over time no damage is found the service intervals can be extended. It should be noted that operation outside of the specifications of the probe will result in voiding Hach factory warranty.