How much carbon dioxide (CO2) is used by the Orbisphere 6110, and will it be released into the atmosphere?

Dokument-ID

Dokument-ID TE9623

Veröffentlichungsdatum

Veröffentlichungsdatum 08.05.2018
Frage
How much carbon dioxide (CO2) is used by the Orbisphere 6110, and will it be released into the atmosphere?
Zusammenfassung
Orbisphere 6110 CO2 consumption
Antwort
The Orbisphere 6110 is not a closed loop system and will exhaust a small amount of CO2 from the analyzer. The operating purge rate is 0.25 mL/s, with a 5 second cycle full blast every 10 minutes which will equal about 40 L/day of gas.

A 50 L cylinder of liquid CO2 will generate 27000 L of CO2 gas, so you will end up with 650 days of gas supply.