What is the difference between blue method and yellow methods for phosphates?

Dokument-ID

Dokument-ID CO201

Veröffentlichungsdatum

Veröffentlichungsdatum 09.04.2018
Frage
What is the difference between blue method and yellow methods for phosphates?
Zusammenfassung
Blue method vs yellow method for phosphate
Antwort
Reactive phosphorous can be measured colorimetrically using a number of different chemistries.  First the phosphorous reacts with molybdate in an acidic solution to form a phosphomolybdate complex.  For low range measurements the phosphomolybdate complex is reduced with either an amino acid or ascorbic acid, creating a characteristic molybdenum blue species.  High range measurements can be made by reacting the phosphomolybdate complex with vanadium to create a yellow species.  Both the blue and yellow complexes are proportional to the concentration of phosphorous in the sample.