What actions should be taken when a U53 totalizer is showing "+++++"?

Dokument-ID

Dokument-ID TE10015

Veröffentlichungsdatum

Veröffentlichungsdatum 12.06.2018
Frage
What actions should be taken when a U53 totalizer is showing "+++++"?
Zusammenfassung
U53 showing over range
Antwort
When the sc200 shows "+++++" for the totalizer, the totalizer has gotten so high that it is out of range for the display. To allow the sc200 to display the current totalizer, the multiplier needs to be increased. Increasing the totalizer can be done as follows: Main Menu > Volume Format > Set Multiplier. The multiplier should be changed to one unit higher than the current value and then Enter must be pressed in order to save the change.  Upon return to the home screen, the totalizer values should appear. If the over range message (+++++) does not go away, bring the multiplier up another unit until the totalizer value appears.
 