How is a hard reset performed on a FL90X using FSData Desktop?

Dokument-ID

Dokument-ID TE10026

Veröffentlichungsdatum

Veröffentlichungsdatum 04.03.2020
Frage
How is a hard reset performed on a FL90X using FSData Desktop?
Zusammenfassung
Hard reset on a FL90X
Antwort
To perform a hard reset on the FL90X, please follow the instructions below:
  1. Download all data.  For further instructions on downloading data, please see Article Link
  2. Save your program file by going to the instrument dashboard and clicking "Save to File" in the lower right hand corner.
  3. Go to "Instrument Tools" and click on "Update Firmware". Check to make sure the firmware version is the latest version, then click update firmware. This step will take about five minutes.
  4. Click on "Restore". This will restore the logger to factory defaults.  This step will take about two minutes.
  5. Click on "Clear", then select "Clear all". This step will take about two minutes.
  6. Load your program from the Instrument Dashboard by clicking on the "Load from File" button in the lower left hand corner.
  7. Write your program to the logger.