How should the dashes in the Fluoride Interference Correction Table of the Aluminum Method 8326 be interpreted?

Dokument-ID

Dokument-ID TE10680

Veröffentlichungsdatum

Veröffentlichungsdatum 22.10.2019
Frage
How should the dashes in the Fluoride Interference Correction Table of the Aluminum Method 8326 be interpreted?
Zusammenfassung
Aluminum Method 8326 Fluoride interference table dashes
Antwort
For method 8326, the upper limit for Aluminum determination is 0.250 mg/L Al3+. Values given in the correction table that would be above this range are represented by dashes and can only indicate the value would be above the range and not a specific value.