PIB 37-315 New Magnetic Stirrer LYW854

Dokument-ID

Dokument-ID PIB195

Veröffentlichungsdatum

Veröffentlichungsdatum 28.12.2018
Titel
PIB 37-315 New Magnetic Stirrer LYW854
Zusammenfassung
New Magnetic Stirrer LYW854; EU Only; Hach Lab; 12/28/2018
PIB Version
Product Change/Update/Other
Level One - Customers-Distributors-Internal


-Introduction:
A re-designed magnetic stirrer is now in the box of the Hach Lab Accessory Part Number LYW854. This change was necessary because of the obsolescence of the old stirring unit.

The new stirrer features

  • A new housing design
  • The Stirring plate diameter increased by 20mm
  • The picture/logo below the stirrer glass is exchangeable
  • The speed range is variable between 100 and 2500 rpm
  • EU list price changed to 399€

 
Scope of delivery:

  • Magnetic stirrer
  • Plug-in power supply with 5 exchangeable country adapters
  • Printed Multilingual Operating Instruction (29 languages)
  • 5 pre-cut A4 paper sheets


LYW854


 
 
The stirring plate diameter is increased by 20mm and the stirring speed is controlled via touch keys. Moreover, the photo or logo picture below the glass stirring plate can be customized and easily exchanged by the user.
 

 
5 pre-cut A4 paper sheets are included for creation of self-printed pictures.
 

5 country adapters are included.

 
 
Product comparison  Old LYW854
New LYW854IKA myPlate Show product details
Number of stirring positions 1 1
Set-up plate dimensions Ø 115 mm Ø 135 mm
Set-up plate material Glass Glass
Stirring quantity max. per stirring position (H2O) 1 l 1.5 l
Speed display LED LED
Speed control Turning knob Touch keys
Speed min. (0) 100 rpm (0) 100 rpm
Speed max. 2500 rpm 2500 rpm
Direction of rotation right left
Setting accuracy speed 50 rpm 10 rpm
Self-heating of the set-up plate by max. stirring (RT:22°C/duration:1h)   2°K
Stirring bar length max. 30 mm 30 mm
Timer No No
Operation at a terrestrial altitude max. 2000m max. 2000m
Permissible ambient temperature min. 5 °C 5 °C
Permissible ambient temperature max. 40 °C 40 °C
Permissible relative humidity 80 % 80 %
Protection class according to DIN EN 60529 IP 54 IP 54
Power input 100 - 240 V, 50/60 Hz, 4 W 100 - 240 V, 50/60 Hz, 4 W
DC Voltage 12 V= 12 V=
Current consumption 300 mA 300 mA
Width x Depth x Height 145 x 160 x 45 mm³ 152 x 164 x 30 mm³
Weight 0.54 kg 0.5 kg

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.