Do Hach pH meters use automatic temperature compensation?

Dokument-ID

Dokument-ID TE11077

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
Do Hach pH meters use automatic temperature compensation?
Zusammenfassung
Temperature compensation on Hach pH meters.
Antwort
All currently available Hach meters use automatic temperature compensation.

The only exception would be if an analog probe that does not have a built in temperature sensor (such as the Sension+ 5209 Product # LZW5209.97.0002 for example) is used with a SensION+ benchtop meter and a separate temperature probe (such as the Radiometer T203 Product # E51M008 for example) is not being used as well. In this case the meter will compensate to the manually entered temperature which defaults to 25 °C.