What is the required quality of instrument air for oxygen concentrators used in B3500 and B7000 BioTector analyzers?

Dokument-ID

Dokument-ID TE11178

Veröffentlichungsdatum

Veröffentlichungsdatum 19.10.2020
Frage
What is the required quality of instrument air for oxygen concentrators used in B3500 and B7000 BioTector analyzers?
Antwort
Quality of the instrument air for oxygen concentrators used in B3500 and B7000 analyzers should meet ISO 8573-1 Class 1 criteria regarding oil and dust content and Class 3 regarding the water content.
When air filter pack is used compressed air of Class 3 or higher can be used. If worse air class is used then frequent replacement of filter pack components is expected.   