How are indicators delivered from PROGNOSYS™ to the operator or technician?

Dokument-ID

Dokument-ID TE11948

Veröffentlichungsdatum

Veröffentlichungsdatum 09.10.2019
Frage
How are indicators delivered from PROGNOSYS™ to the operator or technician?
Zusammenfassung
Information on how PROGNOSYS™ gets information to it's users.
Antwort
Part of PROGNOSYS™ Frequently Asked Questions

The measurement and service indicators display both visually and numerically through a color-coded display on the sc1000 controller. The information can also be out to a PLC/SCADA system.

For example, the measurement health indicator uses yellow, green and red bars to indicate how confident PROGNOSYS™ is in the measurement value. It's looking at some internals within the sensor, such as voltages or reference potentials, as well as the calibration service record to present a measure of confidence in the reading.

When the reading drops below 100%, the operator can touch the color-coded bar and PROGNOSYS™ will display the necessary maintenance task or the reason why it is below 100% health.

The service indicator is based on timers and internal signals within the sensor itself and alerts the operator about when the next service is due.