How to wire a combination style pH/ORP sensor to an external pH gateway

Dokument-ID

Dokument-ID TE12193

Veröffentlichungsdatum

Veröffentlichungsdatum 01.07.2022
Frage
How to wire a combination style pH/ORP sensor to an external pH gateway
Zusammenfassung
wire combo pH probe to pH gateway.  PC1R1A, PC1R2A, PC1R1N AND 6120600
Antwort
The below information has been put together to assist with the wiring of combination pH/ORP electrodes and the pH/ORP external gateway, PN: 6120600.
If you are using a combination pH probe without temperature compensation, example part number PC1R1N, your sensor will not have the White/'Temp +' or Red/'Temp -' wires. Terminals J3-2 and J3-3 will be left empty.

Pictures of a sensor wired to the gateway - sensor signal without ground rod


Pictures of a sensor wired to the gateway - sensor signal with ground rod.  (Most applications will require wiring with the ground rod signal as seen below)


See the following links/articles for more information on wiring these sensors 
sc100 Combination pH-ORP Analysis System User Manual-Digital PC sc and RC sc ¾-inch
How do you wire a Combination style pH or ORP sensor to a sc-controller?
How are combination probes wired to Analog pH Input Modules (Catalog Number/Product Code 9012900) on the sc200 controller ?








 