How do I configure & verify the SC200 relay outputs?
Antwort
The following are the basic steps for configuring the Alarm Relays on the SC200 controller:

To configure the Alarm Relays:

From the HOME screen, select MENU
  • Select SC200 SETUP
  • Select RELAY SETUP
  • Select from:
    • RELAY A
    • RELAY B
    • RELAY C
    • RELAY D
  • Select SOURCE
  • Select the sensor (by name or serial number) that this alarm should represent, and press ENTER
  • Select SET PARAMETER
  • Select the measured value to be alarmed (example: pH or Temp), and press ENTER
  • Select SET FUNCTION
  • Select ALARM
  • Select ACTIVATION
  • Select LOW ALARM
  • Using the arrow keys, enter the measured value that will be represented by a low alarm signal (example: 4 pH = low alarm), then press ENTER
  • Select HIGH ALARM
  • Using the arrow keys, enter the measured value that will be represented by a high alarm signal (example: 10 pH = high alarm), then press ENTER
  • Select BACK (on the keypad)
  • Select SET TRANSFER
  • Select ACTIVE
Alarm Relays are now active.

To verify the Alarm Relay
  • From the HOME screen, select MENU
  • Select TEST/MAINT
  • Select TEST RELAY
  • Select from:
    • RELAY A
    • RELAY B
    • RELAY C
    • RELAY D
  • Select ENERGIZE to pulse an output from the SC200.