What is the air flow path for reagent delivery on a 5500sc or 961Xsc analyzer?

Dokument-ID

Dokument-ID TE12326

Veröffentlichungsdatum

Veröffentlichungsdatum 1581525599000
Frage
What is the air flow path for reagent delivery on a 5500sc or 961Xsc analyzer?
Zusammenfassung
Image and description showing main components for air flow path for reagent delivery on a 5500sc or 961Xsc.
Antwort
The labeled image below shows the main components of the reagent air supply and flow indicators. A system which is working correctly will regulate the reagent air pressure from 3.95 to 4.10 psi and typically require the air pump to only come on a couple seconds every one to two hours. It is important for all fittings to be tight and bottle caps snug.





See the following articles for more information on reagent delivery and troubleshooting information: