How do I replace the cap on the LDO 2 sc sensor and what do I do with the calibration cap?

Dokument-ID

Dokument-ID TE10890

Veröffentlichungsdatum

Veröffentlichungsdatum 03.04.2022
Frage
How do I replace the cap on the LDO 2 sc sensor and what do I do with the calibration cap?
Antwort
  1. Ensure that the probe is connected and being supplied with power.
  2. Clean the sensor.
  3. Remove and discard the old sensor cap.
  4. Install the sensor setup cap (calibration cap).
  5. Press the button on the setup cap. The LED lights up for 45 seconds and then goes into a sequence of flashes.
    • Single flash - 10 times, the setup was successful. Go to step 6.
    • Double flash - 38 times, remove the sensor setup cap. Repeat steps 4 and 5.
  6. Remove the sensor setup cap. Store the sensor setup cap in a dry place.
  7. Install the new sensor cap.
Validate the sensor cap code. To do this, open the menu overview in the controller.
Open the SENSOR SETUP menu and select the correct sensor.
Then open DIAG/TEST and SENSORINFO.
Ensure that the cap label on the display matches the label on the cap.