How is the zero calibration enabled on a 5500sc/ 9610sc Silica analyzer

Dokument-ID

Dokument-ID TE12447

Veröffentlichungsdatum

Veröffentlichungsdatum 15.04.2020
Frage
How is the zero calibration enabled on a 5500sc/ 9610sc Silica analyzer
Zusammenfassung
5500sc/ 9610sc Silica analyzer Zero Cal
Antwort
Any software version greater than 1.09 will have the option to perform a zero calibration in addition to the slope calibration. 
The zero calibration option can be enabled/disabled in the calibration menu.
This information can be found in the calibration section of the full user manual here
To initiate a calibration immedately: 
  1. Press cal to display the calibration menu.
  2. Select START MANUAL CAL
  3. Select ZERO CAL if initiating a calibration immediately
To schedule a future calibration 
  1. Press cal to display the calibration menu.
  2. Select SET AUTO CALIBRATION 
  3. Select ZERO CAL
Please note: if performing a zero calibration, change the blank value in the menu > REAGENTS/STANDARDS. > ENTER BLANK VALUE  to zero instead of what is written on the molybdate bottle. This allows the analyzer to adjust the point as needed through the zero calibration operation.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.