Lachat Method: Ammonia 10-107-06-1-B
Zusammenfassung
Method # |
Range |
MDL |
Units |
Matrix |
Rev. Date |
10-107-06-1-B |
0.05-5.0 |
0.007 |
mg/L NH3-N |
Waters |
27-Aug-01 |
Verfahrens-Dokument
This method is based on the Berthelot reaction. Ammonia reacts with alkaline phenol, then with sodium hypochlorite to form indophenol blue. Sodium nitroprusside (nitroferricyanide) is added to enhance sensitivity. The absorbance of the reaction product is measured at 630 nm, and is directly proportional to the original ammonia concentration in the sample.
*Please note diagrams are not included. Please contact Lachat Technical Support or reach out to your Lachat sales representative for a version containing the method diagram.