Lachat Method: Ammonia 10-107-06-3-D

Dokument-ID

Dokument-ID ME140

Veröffentlichungsdatum

Veröffentlichungsdatum 18.01.2022
Titel
Lachat Method: Ammonia 10-107-06-3-D
Zusammenfassung
Method # Range MDL Units Matrix Rev. Date
10-107-06-3-D 0.005-0.25 0.001 mg/L NH3-N Waters 26-Aug-03
Verfahrens-Dokument
Ammonia in the sample reacts with hypochlorite ions which are generated in situ by alkaline hydrolysis of sodium dicholorisocyanurate. This reaction forms monochloramine which then reacts with salicylate ions in the presence of sodium nitroprusside. A blue, indophenol-type compound is formed which is measured colorimetrically at 660 nm.

*Please note diagrams are not included.  Please contact Lachat Technical Support or reach out to your Lachat sales representative for a version containing the method diagram.