Lachat Method: Ammonia 12-107-06-2-H

Dokument-ID

Dokument-ID ME151

Veröffentlichungsdatum

Veröffentlichungsdatum 18.01.2022
Titel
Lachat Method: Ammonia 12-107-06-2-H
Zusammenfassung
Method # Range MDL Units Matrix Rev. Date
12-107-06-2-H 1-100 0.06 mg/L NH3-N Air Monitoring (above soil) 18-Mar-13
Verfahrens-Dokument
Volatile ammonia is trapped (from the air) using sponges saturated with a glycerol/phosphoric acid solution contained in sponges. The sponges are extracted with water, for a final solution with a concentration of 0.029M glycerol/0.18M H3PO4. The glycerol/ H3PO4 extract is analyzed for ammonia by the salicylate method. When ammonia is heated with salicylate and hypochlorite in an alkaline phosphate buffer an emerald green color is produced which is proportional to the ammonia in the solution. The color is intensified by adding sodium nitroprusside.

*Please note diagrams are not included.  Please contact Lachat Technical Support or reach out to your Lachat sales representative for a version containing the method diagram.