Lachat Method: Ammonia 18-107-06-1-A

Dokument-ID

Dokument-ID ME160

Veröffentlichungsdatum

Veröffentlichungsdatum 18.01.2022
Titel
Lachat Method: Ammonia 18-107-06-1-A
Zusammenfassung
Method # Range MDL Units Matrix Rev. Date
18-107-06-1-A 1.75-140 0.08 mg/L NH3-N Aqueous Formulations 10-Aug-09
Verfahrens-Dokument
This method is based on the Berthelot reaction. Ammonia reacts with alkaline phenol, then with sodium hypochlorite to form indophenol blue. Sodium nitroprusside (nitroferricyanide) is added to enhance sensitivity. The absorbance of the reaction product is measured at 630 nm, and is directly proportional to the original concentration.

*Please note diagrams are not included.  Please contact Lachat Technical Support or reach out to your Lachat sales representative for a version containing the method diagram.