Lachat Method: Calcium 12-120-02-1-B

Dokument-ID

Dokument-ID ME174

Veröffentlichungsdatum

Veröffentlichungsdatum 11.05.2020
Titel
Lachat Method: Calcium 12-120-02-1-B
Zusammenfassung
Method # Range MDL Units Matrix Rev. Date
12-120-02-1-B 0.25-50.0
10-1000
0.05
0.7
mg/L Ca Soil Extracts 16-May-12
Verfahrens-Dokument
Soil extracts are injected into a 1M Ammonium acetate carrier and mixed with an acidic solution of o-cresolphthalein complexone and 8-hydroxyquinoline. Calcium is complexed by the ocresolphthalein and magnesium interference is eliminated by adding 8-hydroxyquinoline as a magnesium chelating agent. The pH is then raised to about 10 and the purple colored complex produced is proportional to the calcium concentration.

Download the method

*Please note diagrams are not included.  Please contact Lachat Technical Support or reach out to your Lachat sales representative for a version containing the method diagram.