How can I transfer my output signal when there is an error on an APA6000?

Dokument-ID

Dokument-ID TE12787

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
How can I transfer my output signal when there is an error on an APA6000?
Zusammenfassung
APA6000 transfer recorder output signal upon analyzer error
Antwort
The default setting for the recorder output on an APA6000 is HOLD when the analyzer has an error or measurement failure.
If an operator would prefer the recorder output to transfer to a maximum or minimum value upon error or measurement failure, the below setting can be changed to either 'Go Max' or 'Go Min'.  See the menu path for this change below, push ENTER after each of the below steps.
Menu > Sensor Menu > Select sensor by name > Recorder Setup > Select RCA or RCB > MEAS FAILURE > choose 'Go Max' if you'd like the output to go to 20mA and 'Go Min' if you'd like the output to go to a 4mA signal.
The below table shows the RCA and RCB recorder setting options.

 