Modified Standpipe for CL17sc to address false sample leak errors

Dokument-ID

Dokument-ID TE12849

Veröffentlichungsdatum

Veröffentlichungsdatum 04.01.2022
Frage
Modified Standpipe for CL17sc to address false sample leak errors
Zusammenfassung
CL17sc modified standpipe for conditioning kit
Antwort
CL17sc may yield false sample leak errors if there is vibration in the sample line, many times caused from a sample pump. A modified standpipe install kit, part number 8576000, was created as a means of lessening the effect of vibration on the sample line and minimizing any false sample leak errors caused by vibrations.
See the below document with information on a modification made to the CL17sc sample conditioning kit standpipe. 

See the following articles for more information on CL17sc sample leaks
How to manually measure sample flow on a CL17sc when getting a LOW SAMPLE FLOW warning