How are Saybolt measurements above 30 displayed on a LICO colorimeter?

Dokument-ID

Dokument-ID TE12882

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
How are Saybolt measurements above 30 displayed on a LICO colorimeter?
Zusammenfassung
Saybolt measurement results
Antwort
On the Saybolt Color Scale, the faintest coloration is Saybolt-color number +30, while the strongest Saybolt-color number is -16.
 
As a result, Saybolt measurements that are above 30 (considered under range for this test), would be displayed as 30, but with a red bar off to the side.
 
Results, that are at 30 (not above), would be displayed as 30, but with a blue bar off to the side.