Can the CL10sc be used to measure chlorine in seawater?

Dokument-ID

Dokument-ID TE12922

Veröffentlichungsdatum

Veröffentlichungsdatum 02.02.2022
Frage
Can the CL10sc be used to measure chlorine in seawater?
Zusammenfassung
using the CL10sc with seawater
Antwort
The CL10 sc will not work well with seawater applications or with any process water that has high salt concentrations, we recommend customers use the CL17 for seawater applications.  There are two main reasons for this recommendation.  The first reason is the  auxiliary electrode in the CL10 sensor is made from stainless steel.  High salt content will lead to corrosion of this electrode which will lead to frequent electrode replacements.  The second reason has to do with the electrolyte used inside the sensor.  High ionic strength of the sample requires special electrolyte which may affect the readings resulting in drift, high swings, electrolyte gets worn out fast, etc. 

See the following articles for more information on chlorine measurement in seawater

Can the CL17 Chlorine Analyzer be used with seawater samples?
Does chloride interfere with the DPD chemistry?
Should I measure free or total chlorine with a seawater application?