Chemistry Explained: Fluoride

Dokument-ID

Dokument-ID TE12980

Veröffentlichungsdatum

Veröffentlichungsdatum 25.05.2022
Frage
Chemistry Explained: Fluoride
Zusammenfassung
Explanation of the reactions and methods for the parameter Fluoride.
Antwort

Introduction:

Fluoride occurs naturally in some ground waters, and a 1-mg/L level is normally maintained in public drinking water supplies for the prevention of dental cavities. Excessive amounts of fluoride cause an objectionable discoloration of tooth enamel called “mottling”. For this reason, a permissible level in drinking water has been established by the USEPA in accordance with the Safe Drinking Water Act.

Chemical Reactions:

SPADNS method

The fluoride analysis involves the reaction of fluoride with a dark red zirconium-dye complex. Fluoride combines with part of the zirconium to form a colorless zirconium-fluoride complex with the net effect of bleaching the color. Measurement of the decrease in color intensity provides an accurate determination of the fluoride concentration. The SPADNS Method is the preferable colorimetric method due to its rapid reaction with fluoride and the stability of the SPADNS reagent.

SPADNS 2 method

Sodium Arsenite is used in the SPADNS method as a reducing agent to prevent interference from chlorine and other oxidants that are typically present in drinking water. The SPADNS 2 test eliminates arsenic from the original SPADNS formulation by using a non-toxic proprietary reducing agent to achieve identical results and test performance. All other chemistry remains the same as the SPADNS method.

Ion-Selective electrode method

The Ion-Selective electrode method requires a Hach ISE Meter and an electrode system consisting of a silver/silver chloride reference electrode and a standard fluoride ion selective electrode. Fluoride measurement is accomplished when a voltage potential is established across the lanthanum fluoride crystal on the end of the electrode; this potential is in direct proportion to the fluoride concentration of the sample. The meter is calibrated with fluoride standards bracketing the expected range. The concentration may be read directly from the meter. A total ionic strength adjustment buffer (TISAB) is used to eliminate interferences in the test, to adjust the pH to an optimum value and to introduce sufficient sodium chloride to mask variations in ionic strength. TISAB reagent uses sodium 1,2-cyclohexanediaminetetraacetic acid (CDTA) for chelation of interfering metals, such as Al3+ and Fe3+, as well as other complexing and buffering agents.

This information is from the document Fluoride for Water and Seawater

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.