Chemistry Explained: Manganese

Dokument-ID

Dokument-ID TE12986

Veröffentlichungsdatum

Veröffentlichungsdatum 12.05.2022
Frage
Chemistry Explained: Manganese
Zusammenfassung
Explanation of the reactions and methods for the parameter Manganese.
Antwort

Introduction:

Manganese is present in ground waters primarily as the divalent ion (Mn2+), due to the lack of subsurface oxygen. Surface waters may contain combinations of manganese in various oxidation states as soluble complexes, or as suspended particles.

The occurrence of manganese in public water supplies presents more of an economic problem than a potential health hazard. Manganese causes dark stains in laundry and on plumbing fixtures, tends to deposit in water lines, and imparts an objectionable taste to beverages such as coffee and tea. Manganese levels in natural waters rarely exceed 1 mg/L, but levels of 0.1 mg/L are sufficient to cause the taste and staining problems. The recommended allowable manganese level in public water supplies is 0.05 mg/L.

Two methods for manganese determination are used in test procedures. The Periodate Oxidation Method gives a simple, rapid test for high levels of manganese. The 1-(2-Pyridylazo)-2-Napthol (PAN) Method is a sensitive, rapid procedure for low levels of manganese.

Chemical Reactions:

Periodate oxidation method

Manganese is oxidized to permanganate using periodate in a slightly acidified water sample. No indicator is necessary. Intensity of the purple color of the permanganate ion is a direct indication of the amount of manganese present in the sample.

3H2O + 2Mn2+ + 5IO4- → 2MnO4- + 5IO3- + 6H+

PAN method

The PAN method employs an Alkaline-Cyanide Reagent. PAN Indicator is added and forms an orange-red colored complex with the manganese ion.

This information is from the document Manganese for Water and Wastewater

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.